NIDMM_ATTR_SAMPLE_INTERVAL

NI-DMM C/CVI/VB

NIDMM_ATTR_SAMPLE_INTERVAL

Specific Attribute

Data
type
Access Coercion High-Level Function
ViReal64 R/W None niDMM_ConfigureMultiPoint

Description

Specifies the amount of time in seconds the DMM waits between measurements. This attribute only applies when the NIDMM_ATTR_SAMPLE_TRIGGER attribute is set to INTERVAL. The default value (–1) ensures that the DMM settles for a recommended time, which is the same as using an immediate trigger.

The NI 4065 and NI 4070/4071/4072 use the value specified in this attribute as additional delay. On the NI 4065 and NI 4070/4071/4072, the onboard timing resolution is 34.72 ns and the valid range is 0–149 s.

On the NI 4060, the value for this attribute is used as the settling time. When this attribute is set to 0, the NI 4060 does not settle between measurements. The onboard timing resolution is 1 µs on the NI 4060.

Only positive values are valid when setting the sample interval.

Note Note  The NI 4050 is not supported.